T1 independent, T2* corrected chemical shift based fat–water separation with multi‐peak fat spectral modeling is an accurate and precise measure of hepatic …

CDG Hines, A Frydrychowicz… - Journal of magnetic …, 2011 - Wiley Online Library
CDG Hines, A Frydrychowicz, G Hamilton, DL Tudorascu, KK Vigen, H Yu, CA McKenzie
Journal of magnetic resonance imaging, 2011Wiley Online Library
Purpose: To determine the precision and accuracy of hepatic fat‐fraction measured with a
chemical shift‐based MRI fat‐water separation method, using single‐voxel MR spectroscopy
(MRS) as a reference standard. Materials and Methods: In 42 patients, two repeated
measurements were made using a T1‐independent, T‐corrected chemical shift‐based fat‐
water separation method with multi‐peak spectral modeling of fat, and T2‐corrected single
voxel MR spectroscopy. Precision was assessed through calculation of Bland‐Altman plots …
Purpose
To determine the precision and accuracy of hepatic fat‐fraction measured with a chemical shift‐based MRI fat‐water separation method, using single‐voxel MR spectroscopy (MRS) as a reference standard.
Materials and Methods
In 42 patients, two repeated measurements were made using a T1‐independent, T‐corrected chemical shift‐based fat‐water separation method with multi‐peak spectral modeling of fat, and T2‐corrected single voxel MR spectroscopy. Precision was assessed through calculation of Bland‐Altman plots and concordance correlation intervals. Accuracy was assessed through linear regression between MRI and MRS. Sensitivity and specificity of MRI fat‐fractions for diagnosis of steatosis using MRS as a reference standard were also calculated.
Results
Statistical analysis demonstrated excellent precision of MRI and MRS fat‐fractions, indicated by 95% confidence intervals (units of absolute percent) of [−2.66%,2.64%] for single MRI ROI measurements, [−0.81%,0.80%] for averaged MRI ROI, and [−2.70%,2.87%] for single‐voxel MRS. Linear regression between MRI and MRS indicated that the MRI method is highly accurate. Sensitivity and specificity for detection of steatosis using averaged MRI ROI were 100% and 94%, respectively. The relationship between hepatic fat‐fraction and body mass index was examined.
Conclusion
Fat‐fraction measured with T1‐independent T‐corrected MRI and multi‐peak spectral modeling of fat is a highly precise and accurate method of quantifying hepatic steatosis. J. Magn. Reson. Imaging 2011;33:873–881. © 2011 Wiley‐Liss, Inc.