Intra‐and interscanner reliability and reproducibility of 3D whole‐brain pseudo‐continuous arterial spin‐labeling MR perfusion at 3T

B Wu, X Lou, X Wu, L Ma - Journal of Magnetic Resonance …, 2014 - Wiley Online Library
B Wu, X Lou, X Wu, L Ma
Journal of Magnetic Resonance Imaging, 2014Wiley Online Library
Purpose To assess the reliability and reproducibility of pseudo‐continuous arterial spin
labeling (pCASL). Intra‐and interscanner reliability and reproducibility are essential before
pooling perfusion imaging data. Materials and Methods Eight volunteers were scanned
three times on two 3.0 T magnetic resonance imaging (MRI) scanners using the pCASL
technique. The first and third tests were conducted on scanner one, while the second test
was on scanner two. The pCASL was acquired with postlabeling delay time (PLD) of 1.5 and …
Purpose
To assess the reliability and reproducibility of pseudo‐continuous arterial spin labeling (pCASL). Intra‐ and interscanner reliability and reproducibility are essential before pooling perfusion imaging data.
Materials and Methods
Eight volunteers were scanned three times on two 3.0T magnetic resonance imaging (MRI) scanners using the pCASL technique. The first and third tests were conducted on scanner one, while the second test was on scanner two. The pCASL was acquired with postlabeling delay time (PLD) of 1.5 and 2.5 seconds. The cerebral blood flow (CBF) of gray matter, white matter, frontal lobe, parietal lobe, occipital lobe, temporal lobe, putamen, posterior cingulate, and thalamus were extracted for comparison. Reliability and reproducibility were measured using the intraclass correlation coefficient (ICC) and within‐subject coefficients of variation (wsCV), respectively.
Results
The pCASL consistently had high intra‐ and interscanner measurement reliability and reproducibility. Reproducibility of the CBF values was higher, with PLD 2.5 seconds than with PLD 1.5 seconds. Different PLDs had considerable influence on the reproducibility of the different regions of the brain.
Conclusion
Although excellent intra‐ and interscanner reliability and reproducibility could support the feasibility of cross‐site pooling of ASL data, pCASL with multiple PLDs may better assess the CBF of the human brain. J. Magn. Reson. Imaging 2014;39:402–409. © 2013 Wiley Periodicals, Inc.
Wiley Online Library