A water-filled body plethysmograph was constructed to measure gas exchange in man. As compared to an air-filled plethysmograph, its advantages were greater sensitivity, less thermal drift, and no change from adiabatic to isothermal conditions after a stepwise change of pressure. When five subjects were completely immersed within it and were breathing to the ambient atmosphere, they had a normal heart rate, oxygen consumption, CO2 output, and functional residual capacity. Pulmonary capillary blood flow ([unk]Qc) during and after Valsalva and Mueller maneuvers was calculated from measurements of N2O uptake. Control measurements of [unk]Qc were 2.58 liters/min per m2 at rest and 3.63 liters/min per m2 after moderate exercise. During the Valsalva maneuver at rest (intrapulmonary pressure: 24, SD 3.0, mm Hg), [unk]Qc decreased from a control of 2.58, SD 0.43, liters/min per m2 to 1.62, SD 0.26, liters/min per m2 with a decrease in pulmonary capillary stroke volume from a control of 42.4, SD 8.8, ml/stroke per m2 to 25.2, SD 5.5, ml/stroke per m2. After release of the Valsalva, there was an overshoot in [unk]Qc averaging +0.78, SD 0.41, liter/min per m2 accompanied by a significant increase in heart rate. Similar changes occurred during and after the Valsalva following moderate exercise. During the Mueller maneuver at rest and after exercise, [unk]Qc, heart rate, and central stroke volume did not change significantly.
Yoshikazu Kawakami, Harold A. Menkes, Arthur B. DuBois
Usage data is cumulative from March 2023 through March 2024.
Usage | JCI | PMC |
---|---|---|
Text version | 112 | 0 |
77 | 14 | |
Scanned page | 293 | 6 |
Citation downloads | 15 | 0 |
Totals | 497 | 20 |
Total Views | 517 |
Usage information is collected from two different sources: this site (JCI) and Pubmed Central (PMC). JCI information (compiled daily) shows human readership based on methods we employ to screen out robotic usage. PMC information (aggregated monthly) is also similarly screened of robotic usage.
Various methods are used to distinguish robotic usage. For example, Google automatically scans articles to add to its search index and identifies itself as robotic; other services might not clearly identify themselves as robotic, or they are new or unknown as robotic. Because this activity can be misinterpreted as human readership, data may be re-processed periodically to reflect an improved understanding of robotic activity. Because of these factors, readers should consider usage information illustrative but subject to change.