Abstract

Rapid pull-through pressure profiles of the normal human upper esophageal sphincter (UES) were simultaneously studied with a conventional three-orifice Honeywell solid-state probe, an eight lumen radially perfused (RP) probe, and a circumferentially sensitive (CS) probe designed to measure UES pressure (UESP) without regard to probe orientation. Pressure curves were digitized and analyzed by computer. The Honeywell probe recorded significantly lower peak pressures than the other two methods, and had wide intrasubject pressure variations (average coefficient of variation, 53%). In contrast, UESP measured with the CS probe was constant for each subject (mean peak UESP, 121 mm Hg; average coefficient of variation, 15%). Anteroposterior RP probe UESP were identical to CS probe pressures. Thus, peak perfused anteroposterior UESP correlates with circumferentially measured sphincter squeeze.

Authors

Richard W. Welch, Kenneth Luckmann, Phillip M. Ricks, Samuel T. Drake, George A. Gates

×